Skip to main content

v1.0

Major VersionM

by Garrett Randolph Hinson

  1. Tap the settings button on the left side of the Fuse touch screen.
    • Tap the settings button on the left side of the Fuse touch screen.

    • Tap checklist.

    • The Maintenance Tasks will appear.

    • Selecting a task to complete will give you step-by-step instructions on how to complete the task.

    • The service interval for each task is displayed to the right of it.

    • Ensure you are wearing a pair of nitrile gloves before completing any tasks.

  2. There are 6 maintenance tasks to be completed.
    • There are 6 maintenance tasks to be completed.

    • Clean Air Filter.

    • Replace Air Filter.

    • Vacuum Powder.

    • Clean Upper Window.

    • Clean Camera Window.

    • Clean IR Sensor Cone.

    • If any task is highlighted red, then it has exceeded its service interval and must be completed.

  3. Select Clean Air Filter. Select begin. Open the air filter access door on the right side of the Fuse. Press down on the button to open the door.
    • Select Clean Air Filter. Select begin.

    • Open the air filter access door on the right side of the Fuse.

    • Press down on the button to open the door.

    • Grab the vacuum hose from the post-processing station and attach the brush-head.

    • Turn on the vacuum by pressing the "VAC" button on the Fuse post-processing station.

    • Very carefully vacuum the air filter and surrounding debris from the chamber.

    • Be sure not to forget to vacuum down the access door as well.

    • Select done to complete the task.

  4. Select replace air filter. Select begin. Open the air filter access door on the right side of the Fuse. Clean the air filter. *reference step 3.
    • Select replace air filter. Select begin.

    • Open the air filter access door on the right side of the Fuse.

    • Clean the air filter. *reference step 3.

    • Depress the spring clamp to release the air filter.

    • Pulling on the grey tabs, pull the air filter out.

    • Grabbing a new air filter, replace in the same position as the old one.

    • Replace the spring clamp to lock the new air filter into place.

    • Close air filter door. Select done.

  5. Using the "Unlock Door" button on the Fuse touch screen, open the door. If the build chamber is inserted, remove it and place it in the Fuse post-processing station. Grab the vacuum hose from the post-processing station and attach the brush-head.
    • Using the "Unlock Door" button on the Fuse touch screen, open the door.

    • If the build chamber is inserted, remove it and place it in the Fuse post-processing station.

    • Grab the vacuum hose from the post-processing station and attach the brush-head.

    • Vacuum out any accumulated powder from the bottom, side, and top of the printer.

    • Close door. Select done.

  6. Grab a swab, the mirror, and ethanol from the cabinet below the Fuse. The LEDs inside the Fuse chamber are very bright. Following the guide on the Fuse will automatically dim them for you. Loosen the screw to drop the optical cassette tray and remove the optical cassette. *reference Fuse printing Dozuki.
    • Grab a swab, the mirror, and ethanol from the cabinet below the Fuse.

    • The LEDs inside the Fuse chamber are very bright. Following the guide on the Fuse will automatically dim them for you.

    • Loosen the screw to drop the optical cassette tray and remove the optical cassette. *reference Fuse printing Dozuki.

    • Once the optical cassette is removed, return the tray to the UP position.

    • Using the mirror wand, located the upper window. It will be towards the rear.

    • The mirror wand will have a switch on the back of it to turn on its LEDs.

    • Once you have located the upper window, wet the foam swab with ethanol.

  7. Wipe around the silver side walls  of the upper window in a single direction.
    • Wipe around the silver side walls of the upper window in a single direction.

    • Repeat with a new swab if necessary to remove any streaks.

    • Replace the optical cassette.

    • Close door. Select done.

  8. Grab a swab, the mirror, and ethanol from the cabinet below the Fuse.
    • Grab a swab, the mirror, and ethanol from the cabinet below the Fuse.

    • Using the "Unlock Door" button on the Fuse touch screen, open the Fuse door.

    • Remove the optical cassette and return tray to the UP position.

    • Using the mirror wand, locate the camera window.

    • Wet a foam swab with a small amount of ethanol.

    • Wipe across the lens of the camera in a single direction multiple times.

    • Once the ethanol has evaporated, complete a visual inspection to ensure the lens is clean. Repeat cleaning if necessary.

    • Replace optical cassette. Close door. Select done.

  9. Grab a foam swab and ethanol from the cabinet below the Fuse. Using a step stool, locate the IR sensor access door on top of the Fuse. Pulling on the IR sensor, remove it from its mount and unplug the yellow ethernet cable.
    • Grab a foam swab and ethanol from the cabinet below the Fuse.

    • Using a step stool, locate the IR sensor access door on top of the Fuse.

    • Pulling on the IR sensor, remove it from its mount and unplug the yellow ethernet cable.

    • Set the IR sensor off the side on a flat surface.

    • Wet a foam swab with a small amount of ethanol.

    • Begin wiping the inside of the IR sensor cone.

    • Visually inspect the cone of any excess debris. Repeat cleaning if necessary.

    • Reattach the ethernet cable and replace the IR sensor. It will illuminate green if seated properly.

Finish Line

Garrett Randolph Hinson

Member since: 09/14/2020

250 Reputation

4 Guides authored

Team

TW: Staff Member of TW: Staff

172 Members

279 Guides authored

2 Comments

We should add either another guide or a few extra steps of maintenance that the machine doesn’t tell you about (laurolactum buildup mainly, im sure there are other things we should look out for)

Brandon Charles Hill - Reply

add levelling the printer

Quinn Colville - Reply

Add Comment